A galvanometer of resistance 100 Ω has 50 divisions on its scale and has sensitivity of 20 μA/division. It is to be converted to voltmeter with three ranges, of 0 - 2 V, 0 - 10 V and 0 - 20 V. The appropriate circuit to do so is:
We have a moving-coil galvanometer whose resistance is . The scale has 50 divisions and the current sensitivity is per division.
First we find the full-scale (maximum) current of the galvanometer. Since each division needs and there are 50 divisions, we write
To convert the galvanometer into a voltmeter, we keep the current through it limited to this same while different external voltages are applied. For each desired range we first write Ohm’s law, then solve for the total resistance that must be in series with the galvanometer.
For the 0 - 2 V range, the total resistance required is
This total already includes the galvanometer’s 100 Ω, so the series resistance needed is
For the 0 - 10 V range, the total resistance required is
and therefore the series resistance for this range alone must satisfy
For the 0 - 20 V range, the total resistance required is
so the required series resistance is
In a multirange voltmeter the usual practice is to place resistors in series such that:
Let these resistors be connected sequentially with closest to the galvanometer. Then we require
This gives the set with the ranges obtained in the order 2 V, 10 V, 20 V as required.
Comparing these values with the options, we see that they match exactly with Option D.
Hence, the correct answer is Option D.